Gas chromatography mass spectrometry GC-MS is a high precision Gas chromatograph mass spectrometer with pre-filter mass analyzer and electron multiplier ensuring high sensitivity. Desorption for solids is performed at controlled temperature between 100 ~ 350 °C with a flow rate of 1 to 1.5 ml / min. With seven stage temperature programming and an interface temperature of 450 °C the analysis is sensitive and specific.
SKU Code: LD-GC-MS-879
Vacuum | Turbo molecular pump ( 67 L/s ) |
Detector | High energy dynode electron multiplier |
Flow rate | 1 ~ 1.5 ml / min |
Scan rate | Upto 10000 amu/s |
Stability | ± 0.10 amu / 48 hrs. |
Mass range | 1.5 ~ 1000 amu |
Resolution | Unit resolution |
Split mode | Split / split less |
Sensitivity | Full scan ( S / N is ≥ 30 : 1 ) |
Split ratio | 1000 : 1 |
Auto sampler | Optional |
Heating rate | Upto 120 °C / min |
Pressure range | 0 ~ 100 psi |
Room temperature | 4 °C ~ 450 °C |
Inlet temperature | 450 °C |
Maximum flow rate | 10 ml / min |
Pressure control mode | Electronic pressure control ( Supports CV and CC ) |
Temperature programming | 7 stages / 8 platforms |
Filament emission current | 0 ~ 350 µA |
Ion – source temperature | 100 °C ~ 350 °C |
EI source ionization energy | 5 eV ~ 250 eV |
GC-MS interface temperature | 450 °C |
Mass Spectrometry Specifications | |
Gas Chromatography Specifications |
Used in identifying and quantifying volatile organic compounds in mixtures, testing residual solvents, identification of trace impurities in liquids or gases, evaluating extracts from plastics, contaminants on semiconductor wafers or other technology products |