Laser Particle Size Analyzer LD-LLPA-C11
Laser Particle Size AnalyzerLD-LLPA-C11is an automatic, benchtop unit with compact design, performs both wet and dry dispersion measurements. Based on Mie and Fourier (Fourier and inverse Fourier) optical system, scattering theory with ‘Laser Light diffraction technology’. It offers 0.02 to 2600 μm (Wet dispersion mode) and 0.1 to 2600 μm (Dry dispersion mode) measurement range. Analysis is made easy by incorporating highly advanced automatic SOP (standard operating procedure), to deliver reliable measurement performance, fast and accurate results.
SKU Code: LD-LLPA-C11
Also Available in below capacities
Measuring range : 0.1 μm - 300 μm
Measuring range : 0.01 μm - 2000 μm
Measuring range : 0.01 μm - 780 μm
Measuring range : 0.1 μm - 1000 μm
Measuring range : 0.1 μm - 2000 μm
Measuring range : 0.1 μm - 500 μm
- Specifications
- Product Applications
Size |
Measuring range
Optics |
Red light source
Weight |
23 kg |
Accuracy |
≤ 0.5 % (wet), ≤ 1 % (dry) |
Detector |
Arrangement
Quantity |
92 pieces (forward, lateral, backward) |
Software |
Report
Auto test |
Enable |
Interface |
USB 2.0 interface |
Air filter |
3 µm, 0.3 µm, 0.01 µm |
Technology |
Laser light diffraction |
Data export |
Excel, Pdf, Word, Jpg and etc. |
Lens design |
F-theta lens |
Power supply |
AC 220 V, 50 Hz |
Air flow rate |
0 to 6000 ml/ min |
Repeatability |
≤ 0.5 % (wet), ≤ 1 % (dry) |
Testing speed |
3 kHz |
Air compressor |
Gas container ≥ 60 L, pressure ≤ 5 bar |
Vacuum cleaner |
Wet dust collector or bag filter |
Water capacity |
600 ml |
Analysis theory |
Mie and Fourier scattering theory |
Lens arrangement |
Dual lens on the right and left side of sample cell |
Operating system |
Windows XP, Windows 7, 8 or 10 |
Resolution ratio |
Single peak, Double peak, Multi-peak |
Dispersion system |
Ultra sound 50 W, 38 kHz dry-burn protection system |
Water circulation |
Centrifugal pump 3000 to 8000 ml/ min, auto water intake and rinsing |
Dimension (L×W×H) |
705 × 318 × 295 mm |
Light path adjustment |
Intelligent automatic alignment |
Effective focal length |
223 mm |
Number of size classes |
More than 100 customized grades |
Hardware specifications |
Intel core I5, 4 GB RAM, 250 GB HD |
Sample dispersion system |
Dispersion type
Typical measurement time |
≤ 10 s |
Particle size distribution |
Suspension, Emulsion, Dry powders |
Used in wet and dry particle size analysis across pharmaceutical, soil and cement, food environment, life science and other industries. |
|
|
|
|
|