Nano spectrophotometer is micro volume UV spectrophotometer which takes measurements at 230 nm, 260 nm, 280 nm. It utilizes CCD detector, Xenon flash lamp as light source which has high stability and long operating life. They are used to detect micro volume quantities of nucleic acids and other concentrated or dilute components is specially designed to analyze the concentration of pure of nucleic acid. The data can be printed with built in printer and can be transferred via SD- RAM card or USB. It requires only minimal amount of sample for accurate measurement.
SKU Code: LD-LNS-B10
Weight | 2.8 kg |
Dimension | 210 x 268 x 181 mm |
Data Output | USB, SD-RAM Card |
Path length | 0.2 mm (For high concentration measurement); 1.0 mm (For ordinary) |
Light source | Xenon flash lamp |
Absorbance range | 0.02 - 80 A (10mm equivalent) |
Measurement time | < 8 s |
Wavelength range | 230 nm, 260 nm, 280 nm |
Operating Voltage | 24 V DC |
Absorbance accuracy | 1 % (7.332 Abs at 260 nm) |
Minimum Sample size | 0.5 - 2.0 µl |
Absorbance precision | 0.003 Abs |
Software Compatibility | Win CE |
Detects protein up to | 0.1 - 100 mg/ml |
Sample Pedestal Material | Aluminum alloy and Quartz fiber |
Standby Power Consumption | 5 W |
Detects nucleic acid up to | 10 - 4000 ng/µL (dsDNA) |
Operating Power Consumption | 40 W |
The nano spectrophotometer is used in laboratory, food industries, chemistry, microbiology, medicine research and development. |