Real Time PCRLD-LRTP-A20is a thermo-cycler having 48 wells of sample capacity and require only 0.2 ml volume sample tubes. Features four detection channels, three block (16 tubes each block) configuration, alarm system, automatically controlled hot-lid for uniform pressure and heat preservation, compact design. Advanced temperature control system, powerful and user-friendly software system, novel fluorescence detection technology etc. ensures relative and complete quantification, precise differentiation and highly sensitive results in a single step procedure. It is used in diagnostic labs, molecular biology, genetic engineering, archaeology, forensics, etc.
SKU Code: LD-LRTP-A20
Socket | RS232, USB, Bluetooth |
Weight | 12 kg |
Programs | Absolute Quantification, Relative Quantification, SNP Analysis, Melting Curve Genotyping, Gradient, HRM, Multi-channel Crosstalk Correction, Background Correction, Automatic Gain, Customized Parameters. |
Dimension | 384 × 353 × 348 mm |
Power Supply | 100 to 240V, 50/60 Hz, 600W |
Dynamics Range | 1-1010copies |
Sample Capacity | 3 × 16 × 0.2 ml (Single tube or Strips 8 CPR Tubes) |
Operation System | Windows XP / Windows Vista /Windows7 / Windows8 |
Device Requirements | Surface Pro series, PC, laptop |
Emission Wavelength | 300 to 800 nm |
Heating/Cooling Rate | 4.0°C /s (max) |
Temperature Accuracy | ≤ ± 0.2°C (full-range), (55°C typical value ≤ ± 0.1°C) |
Excitation Wavelength | 500 to 800 nm |
Temperature Uniformity | ≤ ± 0.4°C (full-range), (55°C typical value ≤ ± 0.3°C) |
Block Temperature Range | 4 to 105°C (Minimum division: 0.1°C) |
Temperature Fluctuation | ≤ ± 0.1°C (full-range), (55°C typical value ≤ ± 0.1°C) |
Hot-lid Temperature Range | 70 to 110°C (adjustable, default: 105°C) |
Ambient Working Conditions | Environmental temperature: 10°C to 30°C Environmental RH: ≤ 70 % Altitude: < 2000 m |
Temperature Range of Hot-lid Working | 70 to 110°C (adjustable, default 105°C) |
Repeatability of Fluorescent Intensity Detection | 5% |
It is used for parallel amplification and quantification of fluorescent tagged DNA or RNA molecules. Widely used to diagnose genetic disease, detect infections in humans and environment, DNA profiling, research etc. |