Biosafety Cabinet Class II A2LD-LBS2-A22is built in with foot switch to avoid airflow turbulence caused by arm movement, motorized front window of two layered laminated toughened glass and anti UV, UV lamp for most efficient decontamination and remote control with all functions for easier and convenient operation. Offers maximum opening of 500 mm (20 inch) and tested opening of safety height 200 mm (8 inch). Features airflow system of 30 % air exhaust and 70 % air recirculation and three protection such as sample, environment and operator. Incorporated with interlock function for blower and front window, UV lamp and front window, UV lamp and blower and LED lamp. Designed with UV life indicator and HEPA filters of two, 99.999 % efficient at 0.3 um.
SKU Code: LD-LBS2-A22
Noise | EN12469 ≤ 58 dB / NSF49 ≤ 61 dB |
Caster | Foot master caster |
Display | LCD display: exhaust filter and down flow filter pressure, filter and UV lamp working time, inflow and down flow velocity, filter life, humidity and temperature, system working time etc. |
UV Lamp | 40 W×1 UV timer, UV life indicator, emission of 253.7 nanometres for most efficient decontamination |
Consumption | 900 W |
ULPA Filter | Two, 99.999 % efficient at 0.12 um, filter life indicator |
Front Window | Motorized, two-layer laminated toughened glass ≥ 5 mm, anti UV |
Gross Weight | 316 kg |
Illumination | ≥ 1000 Lux |
Power Supply | AC 220 V ± 10 %, 50/60 hz; 110 V ± 10 %, 60 hz |
Airflow System | 70 % air recirculation, 30 % air exhaust |
Control System | Microprocessor |
Tested Opening | Safety height of 200 mm (8 inch) |
Inflow Velocity | 0.53 ± 0.025 m/s |
Maximum Opening | 500 mm (20 inch) |
Illuminating Lamp | 16 W×2 LED Lamp |
Down Flow Velocity | 0.33 ± 0.025 m/s |
Waterproof Sockets | Two, total load of two sockets: 500 W |
Work Surface Height | 750 mm |
Package Size (W×D×H) | 1630×1000×1820 mm |
Visual and audio alarm | Filter replacement, front window at unsafe height, abnormal airflow velocity |
External Size (W×D×H) | 1500×750×2250 mm |
Internal Size (W×D×H) | 1350×600×660 mm |
Biosafety Cabinet Class II A2 is widely used in pathology, research laboratories, universities, colleges and hospitals. It is suitable to use for microbiology research in absence of volatile, toxic chemicals and radionuclide. |