Features

  • Common Features
  • Stability indicator displays the status of current measurement
  • Measures Conductivity, TDS, Salinity, Resistivity and Temperature of the liquids
  • Automatic temperature compensation for accurate values over the entire range
  • Selectable temperature units (°C or °F) for different application requirements
  • Reset function automatically resumes all the setting back to factory default options
  • Manual and automatic mode of temperature calibration
  • Auto hold feature to lock, record and read stable measurement endpoint
  • Password protection to avoid unauthorized access to settings
  • Expanded memory stores and recalls up to 1000 readings
  • Conductivity / TDS / Salinity / Resistivity Mode
  • 1 to 3 Conductivity push button calibration points
  • 1 Temperature calibration point
  • Can select cell constant (0.1/1/10) and normalization temperature (20°C or 25°C)
  • TDS conversion factor, seawater, practical salinity measurements mode to meet different water quality measurement requirements
  • Linear or pure water compensation improves the measurement accuracy

Benchtop Conductivity Meter LD-LBCM-A15

Benchtop Conductivity Meter LD-LBCM-A15 with 1 to 3 points push button conductivity calibration, automatically recognizes calibration standards. With setup menu, it does setting of 6 parameters calibration points, temperature units and others. 7 inch TFT display shows settings and measurement values. Temperature compensation by manual & automatic method improves the accuracy of conductivity meter.

SKU Code: LD-LBCM-A15

Also Available in below capacities

Conductivity measurement range : 0 - 20.0, 200.0, 2000 µS / cm, 200.0 mS / cm
Conductivity measurement range : 0.000 μS/cm to 3000mS/cm
Conductivity Mode :
Memory Stores Up to 1000 data sets
Output USB interface
Weight 1.7 kg
Display LCD
TDS Mode
TDS measurement Range0 to 10.00, 100.0, 1000 mg/L, 10.00, 200.0 g/L
TDS Accuracy± 1 % F.S.
TDS Factor0.01 to 1.00 (default 0.5)
TDS Resolution0.01 / 0.1 / 1 or automatic
Connector 6-pin mini plug
Power supply DC 12V, Power Adapter
Salinity Mode
Salinity Range0.00 to 80.00 ppt / 0.00 to 42.00 psu / 0.00 to 8.00%
Salinity Resolution0.01
Salinity Accuracy± 1 % F.S.
Reset Function Yes
Resistivity Mode
Resistivity Range0.00 to 30.00 MΩ
Resistivity Resolution0.01 / 0.1 / 1 or automatic
Resistivity Accuracy± 1 % F.S.
Temperature Mode
Temperature range0 to 105°C, 32 to 221°F
Temperature accuracy± 0.5 °C
Temperature resolution± 1 °C
Temperature calibration points1 point
Temperature coefficient0.0 to 10.0 % per °C
Temperature Compensation Range0 to 100°C / 32 to 212°F
Conductivity Mode
Conductivity Range0.01 to 20.00, 200.0, 2000 µS/cm, 20.00, 200.0 mS/cm
Conductivity Accuracy±0.5 % F.S.
Conductivity Resolution0.001/0.01/0.1/1, automatic
Conductivity Calibration Points1 to 3 Points
Conductivity Solutions10, 84, 1413 µS/cm, 12.88, 111.8 mS/ cm
Conductivity Temperature coefficient0.0 to 10.0 %/°C
Compensation coefficientLinear / Non-Liner / Pure Water
Cell Constant2-Pole Electrode K=0.1, 1, 10 or 4-Pole Electrode
Reference temperature20 or 25°C
Data Hold Function Manual or Automatic Endpoint Detection
Stability Criteria Standard or High
Calibration Due Alarm 1 to 31 Days
Dimension (L × W × H) 240 × 220 × 80 mm
Used for quality checking of water on different parameters across research, QC labs, water management plants and others production units.